Pressure Measurement Specification: |
Operating Pressure Ranges: |
Standard Full Scale Ranges can be Specified of -14.7 psig to: +15 psi, +30 psi, 100 psi, or 300 psi Gauge |
Overpressure: |
The Rated Pressure can be Exceeded by 2X Causing Negligible Calibration Change, Subject to a Maximum Safe Working Pressure of 350 psi. |
Pressure Media: |
Fluids Compatible with 316 Stainless Steel, Hastelloy C276, Nylon, Viton and Brass. |
Accuracy: |
Combined Non-Linearity, Hysteresis and Repeatability +/-0.075% F.S. for Positive or Negative Pressures +/-1 Digit. |
Temperature Effects: |
Over 32 to 105 , the Averaged Span Temperature Coefficient will not Exceed 0.007% of Reading/ F |
Digital Indicator: |
Readout & Display: |
+/-99999 Capability, 0.54 LCD Digits with Additional 16 Text Characters. |
Pressure Scale Units: |
24 Scale Units are Provided: psi, lb/ft2, inHg, inH2O, (40 and 68 F) torr, atm, Pa, kPa, MPa, hPa, mbar, bar, kg/cm2, kg/in2, mmHg, cmHg. mHg, mmH2O, cmH2O, mH2O, Special, 2 Units are Presented on the Display. |
Display Overload: |
Instruments are Capable of Nominal 120% F. S. Above this, an Overrange Error Code is Displayed. |
Resolution: |
+/-0.005% F.S. Maximum |
Response: |
Nominal Rate of Three Readings per Second on the Digital Display. |
Zero Control: |
Offset Correction for Pressure and Electrical Scales by Key-Press, Protected Against Accidental Use. |
Calibration Control: |
Front Panel Key-Pad Calibration Controls for Pressure and Electrical Ranges. |
Position Effect: |
Position Effect |
Electrical Specifications: |
Electrical Inputs: |
Voltage Measurement: Range: Accuracy: Temperature Coefficient: |
+/-50V +/-0.15% of Reading +/-0.02 % F.S. +/-0.006% Reading/ F |
Current Measurement: Range: Accuracy: Temperature Coefficient: |
+/-55mA +/-0.075% of Reading +/-0.005 % F.S. +/-0.006% Reading/ F |
Electrical Outputs: |
Voltage Output: |
24V Nominal, Floating, 25mA Maximum |
Current Output: |
Current Output: Loop Check Fixed 12mA Nominal, Source or Sink. |
Power Supply: |
Battery Powered 4 x 1.5V Zinc Carbon "D" Cells, 50 Hours Nominal Continuous Operation with No Load on Output Terminals. Battery Low Condition Indicated by Display Symbol.
Optional Rechargeable Batteries and External Power Adaptor/Charger Unit Available. |
Environmental Specification: |
Temperature: |
Operating: 15 to 120 F Calibrated: 32 to 105 F Storage: -5 to +140 F |
Sealing: |
Sealed to NEMA 4 |
Safety: |
lEC 1010 EMC Immunity to lEC 801-804 Certification: CE Marked |
Physical Specification: |
Weight: |
9 lbs. |
Dimensions: |
12.6" x 7.7" x 5.0" Nominal |
Pressure Connection: |
1/8" NPT Female |
Carrying Case: |
Supplied with the Instrument as Standard and Includes a Complete Set of Accessories. |