Druck DPI 610 Portable Pressure Calibrator

<< Previous in Druck & SI Pressure (GE Sensing) Next in Druck & SI Pressure (GE Sensing) >>
OUR PRICE: Discontinued
Item Number: DPI 610
Manufacturer: Druck (GE)
Manufacturer Part No: DPI 610
Druck DPI 610

Druck DPI 610 Portable Pressure Calibrator

    The Druck 610 is a self contained, battery powered calibrator which contains a pressure generator, fine pressure control, transmitter power supply & mA measurement capability. The 610 is ideal for 4-20mA loop testing and data storage/documetntion. A ramp and step function is ideal for commisioning loops. The rugged, sealed design is ergonomic and allows lead connection and pump operation without blocking the large LCD display. A highly accurate and easy to use calibrator is only part of the solution to improving overall data quality. The DPI 610, with data storage and RS-232 interface, can reduce calibration times and eliminate data recording errors.  The advanced Model 615 allows for full calibration documentation and two way data transfer. The 615 also allows % error calculation on the spot to show the technician when adjustments are required.

Features:
 

  • High Accuracy, 0.025% Full Scale
  • Ranges from -14.7 to 10,000 PSI
  • PSI for Gauge, Asolute of DP Measurement
  • Internal Pneumatic Vacuum/Pressure Source
  • Pneumatic Generation -22" Hg to 300 PSI
  • Internal Hydraulic Source, 0-6000 PSI
  • Measure: mA, V, Switch State, Temperature
  • Source: Pressure, mA Step, mA Ramp, Actual mA
  • Powers Transmitters from 10 to 24 VDC
  • Stores up to 10,000 Readings
  • Second Pressure Port for Remote Sensors
  • Dual Line LCD: Input & Output on One Display
  • Linkpak Software Documents Calibrations
  • Stores up to 20 Screen Displays
  • Operates in 6 Languages, 25 Pressure Scales
  • Measure mA, Volts, Pressure, Switch State
  • Measure Ambient Temperture
  • Source Pressure, mA, Transmitter Power
  • Test Pressure Switches
  • Perform Leak Tests
  • Pressure Transmitter Calibration
  • mA Loop Testing & Troubleshooting
      The DPI 610PC, 610HC & DPI610I include an integral pressure sensor. Select an appropriate range from the table below:
      Pressure
      Range
      Pneumatic
      DPI 610PC
      Hydraulic
      DPI 610HC
      Indicator
      DPI 610I
      Remote
      Option (B1)
      Accuracy
      %F.S.
      1 PSI (-1) G   G G or D 0.025
      2.5 PSI (-2.5) G   G G or D 0.025
      5 PSI (-5) G or A   G or A G,A or D 0.025
      10 PSI (-10) G or A   G or A G,A or D 0.025
      15 PSI (-15) G or A   G or A G,A or D 0.025
      30 PSI (-15) G or A   G or A G,A or D 0.025
      50 PSI (-15) G or A   G or A G,A or D 0.025
      100 PSI (-15) G or A   G or A G,A or D 0.025
      150 PSI (-15) G or A   G or A G,A or D 0.025
      300 PSI (-15) G or A   G or A G,A or D 0.025
      500 PSI (-15)     G or A G,A or D 0.025
      1000 PSI (-15)   G or A G or A G or A 0.025
      2000 PSI   S,G or A S,G or A S,G or A 0.025
      2500 PSI   S,G or A S,G or A   0.025
      3000 PSI   S,G or A S,G or A S,G or A 0.025
      5000 PSI   S,G or A S,G or A S,G or A 0.025
      6000 PSI   S,G or A S,G or A   0.025
      10,000 PSI       SG or A 0.025
      Values in ( ) Indicate Negative Calibration for Gauge
      and Differential Ranges
      A= Absolute, D = Differential, SG = Sealed Gauge, G = Gauge
      Option B1 is a remote sensor which plugs into the DPI 610/615. 

    Technical Specifications:

    Accuracy: ±0.025% of Full Scale
    Span Shift: 0.5%/500 PSI Line or DP Pressure
    Temp Effect: ±0.002% Reading/Deg F
    Overpressure: Safe to 1.5 X Full Scale
    Input:
        Voltage:
        Current:
        Temperature:
        Switch:

    ±50 VDC, ±0.05% Rdg,
    100 µV Resolution
    ±55 mA, ±0.05% Rdg,
    0.001mA Resolution
    15 to 105 F, ±2 Deg F,
    0.1F Resolution
    Open/Closed
    Output:
        Voltage:
        Current:


    10 VDC, ±0.1% or 24 VDC, ±5%
    0 to 24 mA, ±0.05% Rdg,
    0.001mA Resolution
    Pressure Units: 25 Scale Units, One User Defined
    mA Step:
        Function:
        Function:
        Function:
        Function:
        Function:
    Continious Cycle at 10 Sec Intervals
    Linear: 4, 8, 12, 16, 20 mA
    Linear: 0, 5, 10, 15, 20 mA
    Flow: 4, 5, 8, 13, 20 mA
    Flow: 0, 1.25, 5, 11.25, 20 mA
    Valve: 3.8, 4, 4.2, 12, 19, 20, 21 mA
    mA Ramp: Continious, Configure End
    Value, 60 Sec Interval
    Data Logger: Manual or Automatic to 10,000 Values
    Snapshot: Store up to 20 Screens
    PC Interface: RS-232
    Process Functions: Tare, MAX/MIN, Filter, Flow, % Span
    Languages: English, German, French, Italian,
    Portuguese, Spanish
    Power: Auto Off, Auto Backlight Off, Low
    Battery Indicator
    6 "C" Cells, 65 Hour Use, 110
    VAC Power Available
    Display: 2.36" x 2.36" LCD, ±99999,
    2 Readings per Sec.
    Temperature: Operating: 15 to 120 Deg F,
    0 to 90% Humidity
    Sealing: Generally NEMA 12
    Size: 6.6 Lbs, 11.8"x6.7"x5.5"
    Conformity: EN61010, EN500801-1,
    EN50082-1, CE Mark