Hart Scientific Fluke 9009 Dry Block Calibrator
Item Number: 9009-B-156
Manufacturer: Hart Scientific
Manufacturer Part No: 9009-B-156
Product Options
* denotes required field
Hart 9009 Case Color Selection*
You must select an option for 'Hart 9009 Case Color Selection'.
Spare Insert #1 Selection for Hart 9009*
You must select an option for 'Spare Insert #1 Selection for Hart 9009'.
Spare Insert #2 Selection for Hart 9009*
You must select an option for 'Spare Insert #2 Selection for Hart 9009'.
NIST Traceable Calibration
* Whole number only
|
Hart Scientific Fluke 9009-B Dry Block CalibratorHart Scientific 9009-Y Dry Block Calibrator |
|
Hart Scientific Fluke 9009 Hot & Cold Dual Dry Block
Hart's new Model 9009 Industrial Dual-Block Calibrator lets you calibrate at hot and cold temperatures at the same time. The hot block provides temperatures from 50 to 350 C, while the cold block covers the range -15 to 110 C. Each block is controlled by a precision Hart Scientific temperature controller. Each temperature block includes two wells with removable inserts. You can calibrate four probes at once, or you can calibrate two probes at the same time with an external reference thermometer, or you can use the two temperature wells to get quick “zero” and “span” references for transmitter calibrations. The 9009 is housed in a tough Pelican™ case that is both airtight and watertight, weighing only 10 pounds. Inserts are available to accommodate sensors of any size from 1/16" (1.6 mm) to 7/16” (11.1 mm).The 9009 also delivers the performance you expect from a Hart Scientific temperature source. The cold block is calibrated to within 0.2 C with stability of 0.05 C. The hot block's display is accurate to 0.6 C with stability of 0.05 C. Two blocks in one unit, a total range of -15 to 350 C, portability, durability, versatility, performance and automation. Features:
- Two wells in each block (Test 4 Sensors)
- Rugged, lightweight, watertight enclosures
- Calibrate HOT and COLD temperatures at the same time
- Two independently controlled temperature blocks
- Hot block temperature range from 50 to 350 C
- Cold Block covers the range -15 to 110 C
- Stability 0.05 C Well to well uniformity 0.1 C
- Display resolution: 0.1 Degree
- Stabilization times: 8 minutes
- Well Depth: 4" (102mm)
- Size: 7"H x 10.5"W x 9.75"D (178 x 267 x 248 mm)
- Weight: 10lb (4.5 kg)
- Includes NIST Traceable Calibration Certificate
Technical Specifications:
|
HOT BLOCK |
COLD BLOCK |
Range: |
50 to 350 C (122 F to 662 F)
|
-15 to 110 C (5 F to 230 F) (-8 with hot block at 350 C )
|
Accuracy: |
+/- 0.6 C
|
+/- 0.2 C
|
Stability: |
+/- 0.05 C
|
+/- 0.05 C
|
Uniformity: |
+/- 0.1 C
|
+/- 0.1 C
|
Resolution: |
0.1
|
0.1
|
Heating Times: |
10 minutes from 25 C to 350 C
|
15 minutes from 25 C to 110 C
|
Cooling Times |
30 minutes from 350 C to 100 C
|
16 minutes from 25 C to -15 C
|
Stabilization: |
8 minutes
|
8 minutes
|
Well Depth: |
4" (102 mm)
|
4" (102 mm)
|
Inserts: |
Two 1/4” (6.4 mm) and 3/16” (4.8 mm) inserts included; Others Available |
PC Interface: |
RS-232 included with free Interface-it software |
Power: |
115 VAC (±10%), 3 A, or 230 VAC (±10%), 2 A, specify, 50/60 Hz, 250 W |
Size: |
7" H x 10.5" W x 9.75" D (178 x 267 x 248 mm) |
Weight: |
10 lb. (4.5 kg) |
|
|
|