Druck DPI 740 Precision Barometer

<< Previous in Druck & SI Pressure (GE Sensing) Next in Druck & SI Pressure (GE Sensing) >>
OUR PRICE: Discontinued
Item Number: DPI740
Manufacturer: Druck (GE)
Manufacturer Part No: DPI740
Druck DPI-740 Precision Barometer

Druck DPI-740 Precision Barometer    

    The Druck 740 is a portable battery powered precision
barometer, providing outstanding capabilities in a handheld
package. 
The fundamental sensor technology employed within
this instrument is the Druck Resonant Pressure
Transducer (RPT).
This vibrating element sensor not
only provides accuracies of
+/-0.005 inHg but equally
important has a calibration stability  of
better than 100
ppm per year. This instrument can confidently be
used
in both laboratory and field applications as a first-class
barometric reference device, with the recommended
recalibration
period being one year.
A variety of pressure scales is available to
enable the
operator to display the atmospheric pressure in any
desired units or in height equivalent units for altimetry
and
surveying applications. With the serial RS 232
interface, it is
possible to utilize the DPI 740 data as
input  parameters to
computer systems or simply print
out the  displayed values. 
The DPI 740 ABS molded case has been designed for
hand-held
use, either in field applications, or for use on
the bench with its
integral folding stand.

Features:

  • Accuracy +/-0.004 inHg absolute
  • Ranges to 103 inHg Absolute
  • 6 Digit resolution LCD display
  • Multiple pressure scales and
aeronautical units
  • Alkaline or rechargeable batteries
  • Lightweight hand-held rugged design
  • RS 232 output

 

Technical Specifications:

Range: 22-34"Hg Absolute
Optional Ranges: 1 - 38"HgA, 77"HgA, 103"HgA
Overpressure: Up to 118"HgA will not cause a calibration change.
Pressure Media: Compatible with non-corrosive gases.
Insensitive to media density.
Accuracy: +/-0.02% F.S. (Includes All Effects from 15 to 120 F)
Stability: Better than 100 ppm per annum
Display: 999999 capability LCD digits 0.54" high
Display Overload: 110% F.S. for Each Range.
Above This, Error Code Will Flash.
Resolution: 0.0003"Hg Absolute
Response: 2 Readings Per Second
Calibration Controls: Front panel PIN protected, Also via RS-232.
Operating Modes: Local pressure (QFE)
Sea level pressure (QFF) from user set height
Altitude above user set datum
Process Features: Max/Min, Reading Store, Tare and Programmable Filter
Scales: inHg, psi, atm, torr, inH2O, ftH2O (40 and 68 F), mH2O,
cmH2O, mmH2O,mHg, cmHg, mmHg, kg/in2, kg/cm2, bar,
hPa, inPa, kPa, Pa, mbar Altitude values in ft. and m.
Tare available for surveying.
Position Effect: Negligible
Power Supply: 3 each 1.5V Alkaline AA batteries.
20 hours nominal continuous operation.
Optional rechargeable NiCad batteries
Temperature Operating: 15 to 120 F
Storage: -40 to +160F
Sealing: Generally to NEMA 4
Safety: EN61010, EN50081-1, EN50082-1, CE
Size: 7.5" x 3.5" x 1.4", 1.1 Lbs.
Pressure Connection: Accepts 1/4 in. O/D / 5/32 ID tubing
Serial Output: RS 232 via 6-Way LEMO socket