Martel PTC-8001 Temperature Calibrator

<< Previous in Martel Beta Calibrators Next in Martel Beta Calibrators >>
Martel PTC-8001Thermocouple and RTD Calibrator

OUR PRICE: Discontinued, Replaced by Fluke 724
Item Number: PTC-8001
Manufacturer: Martel Electronics (Beta)
Manufacturer Part No: PTC-8001

Martel PTC-8001 Thermocouple and RTD Calibrator

    The Martel PTC-8001 Precision Temperature Calibrator is an ideal device to handle all of your temperature calibration needs in a truely rugged, low cost package. It combines virtually all widely used thermocouples and RTD's in one device-you'll never have to worry about finding a temperature device you can't calibrate. This is especially true when calibrating "smart" or pulsed Current RTD transmitters where many other calibrators fail to work or operate at reduced accuracy specs. The PTC-8001 handles those applications with ease. 

Features: 

  • Thermocouple & RTD Source/Measure in One Unit
  • Works with "Smart" Pulsed Current Transmitters
  • High Accuracy +/-0.4 C J T/C and +/-0.3 C 4W Pt385 RTD 
  • 10 T/C Types and 8 RTD Types Including PT 385 (1000 ohm) and Cu (10 ohm)
  • Direct Keyboard Entry or Scroll of Output
  • NEMA 4 Rated Case, Rugged Design
  • YSI 400 Series Thermistor Support
  • Standard Mini-Plug or Banana Jack Connections for T/C
  • Store up to Nine Setpoints for Each Output Function
  • RS232 Interface for Remote PC Control
      Technical Specifications 
    T/C Ranges J, K, E, R, S, N, B, L, U including - 10 to 70mV range
    RTD Ranges Pt 385 (100, 200, 500, 1000ohms)
    Pt 392, JIS, Ni 120, Cu10, YSI 400
    Ohms Ranges 0 to 400.00 and 400.0 to 4000.0
    Accuracy T/C Type J: +/-.3 C +/-10uV (+/-0.6 C total error)
    RTD PT100: +/-.1 C +/-.075W (+/-0.3 C total error)
    Ohms (400): +/-0.1 Ohm
    Ohms (4000):+/-1.0 Ohm
    RTS IEX-range 0.01mA to 3mA
    RTD Frequency
    Response
    10 mS - works with al pulsed XMTR's
    Oper. Temp. -10 C to 50 C
    Storage Temp. -20 C to 60 C
    Power Supply 4 AA Alkaline Cells, 30 hours typical operating time
    Size 7.5" x 4" 1.5"

    Range (4 Wire) Min Max Accuracy
    Ohms Read (Low) 0.00 400.00 0.1 Ohm
    Ohms Read (High) 400.0 1500.0 0.5 Ohm

    1500.0 4000.0 1.0 Ohm

    Range Min Max Excitation
    Current
    Accuracy
    Ohms Source (Low) 5.00
    5.00
    400.00
    400.00
    0.1 to 0.5mA
    0.5 to 3mA
    0.15 Ohm
    0.1 Ohm
    Ohms Source (High) 400.0
    1500.0
    1500.0
    4000.0
    0.05 to 0.8mA
    0.05 to 0.4mA
    0.5 Ohm
    1.0 Ohm

    RTD Read and Source
    Range Min Max Accuracy
    Ni120 (672) -80.0 260.0 0.2 C
    Pt100 (385) -200.0 800.0 0.33 C
    Pt100(3926) -200.0 630.0 0.3 C
    Pt100(3916) -200.0 630.0 0.3 C
    Pt200(385) -200.0 630.0 0.8 C
    Pt500(385) -200.0
    500.0
    500.0
    630.0
    0.3 C
    0.4 C
    Pt1000(385) -200.0
    100.0
    100.0
    630.0
    0.2 C
    0.3 C
    Cu10 -100.0 250.0 2.2 C
    YSI400 15.00 50.00 0.05 C

    Thermocouple Read and Source(All Errors Included)
    Range Min Max Accuracy
    J Thermocouple -200.0
    0.0
    0.0
    1200.0
    0.6 C
    0.4 C
    K Thermocouple -200.0
    0.0
    0.0
    1370.0
    0.8 C
    0.5 C
    T Thermocouple -200.0
    0.0
    0.0
    400.0
    0.8 C
    0.5 C
    E Thermocouple -200.0
    0.0
    0.0
    950.0
    0.5 C
    0.4 C
    R Thermocouple -20
    0
    500
    0.0
    500
    1750
    2.4 C
    1.7 C
    1.3 C
    S Thermocouple -20
    0
    500
    0
    500
    1750
    2.4 C
    1.7 C
    1.4 C
    B Thermocouple 600
    800
    1000
    800
    1000
    1800
    2.1 C
    1.7 C
    1.3 C
    L Thermocouple -200
    0.0
    0.0
    900.0
    0.45 C
    0.4 C
    U Thermocouple -200
    0.0
    0.0
    400.0
    0.7 C
    0.45 C
    N Thermocouple -200
    0.0
    0.0
    1300.0
    1.1 C
    0.6 C
    mV Read/Source -10.00 75.00 0.015% +/- 2