Hart Scientific 9143 Field Metrology Well
Item Number: 9143
Manufacturer: Hart Scientific
Manufacturer Part No: 9143
Product Options
* denotes required field
Hart Scientific 9143 Model Options*
You must select an option for 'Hart Scientific 9143 Model Options'.
Hart Scientific 9143 Insert Selection
NIST Traceable Calibration*
* Whole number only
|
|
Hart Scientific 9143 Field Metrology Well
The new 9143 Series Field Metrology Wells extend high performance to the industrial process environment by maximizing portability, speed, and functionality with little compromise to metrology performance. These industrial temperature loop calibrators are perfect for performing transmitter loop calibrations, comparison calibrations, or simple checks of thermocouple sensors. With the “process” option, there is no need to carry additional tools into the field. This optional built-in two-channel readout reads resistance, voltage, and 4–20 mA current with 24 volt loop power. It also has on-board automation and documentation.
Features:
- Lightweight, portable, and fast
- Cool to:
360 C to 33 C in 32 minutes 360 C to 100 C in 14 minutes
- Heat to:
33 C to 350 C in 5 minutes
- Built-in two-channel readout for PRT, RTD, thermocouple, 4-20 mA current
- True reference thermometry with accuracy to +/-0.01 C
- On-board automation and documentation
- Metrology performance in accuracy, stability, uniformity, and loading
Specifications:
Temperature Range at 23 C |
33 C to 350 C(91 F to 662 F) |
Display Accuracy |
+/- 0.2 C Full Range |
Stability |
+/- 0.02 C at 33 C +/- 0.02 C at 200 C +/- 0.03 C at 350 C |
Axial Uniformity at 40 mm (1.6 in) |
+/- 0.04 C at 33 C +/- 0.1 C at 200 C +/- 0.2 C at 350 C |
Radial Uniformity |
+/- 0.01 C at 33 C +/- 0.015 C at 200 C +/- 0.02 C at 350 C |
Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes) |
+/- 0.015 C Full Range |
Hysteresis |
0.03 |
Operating Conditions |
0 C to 50 C, 0 % to 90 % RH (non-condensing) |
Environmental Conditions (for all specifications except temperature range) |
13 C to 33 C |
Immersion (Well) Depth |
150 mm (5.9 in) |
Insert OD |
25.3 mm (1.00 in) |
Heating Time |
5 min: 33 C to 350 C |
Cooling Time |
32 min: 350 C to 33 C 14 min: 350 C to 100 C |
Resolution |
0.01 Deg |
Display |
LCD, C or F user-selectable |
Size (H x W x D) |
290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in) |
Weight |
7.3 kg (16 lb) |
Power Requirements |
100 V to 115 V (+/- 10 %), 50/60 Hz, 1400 W 230 V (+/- 10 %), 50/60 Hz, |
Computer Interface |
RS-232 and 9930 Interface-it control software included |
9143 Includes:
- Users Manual
- Technical Manual on CD
- Serial Cable
- 9930, Interface-itSoftware
- Report of Calibration
- Insert Removal Tool
9143-_-P Includes:
- Users Manual
- Technical Manual on CD
- Serial Cable
- 9930, Interface-it Software
- Report of Calibration
- Test leads
- Insert Removal Tool
- 6-pin DIN Connector for Reference Probe
|
Field Metrology Wells are packed with functionality and are remarkably easy to use. They are lightweight, small, and quick to reach temperature set points, yet they are stable, uniform, and accurate.
|
Field Metrology Wells are packed with functionality and are remarkably easy to use. They are lightweight, small, and quick to reach temperature set points, yet they are stable, uniform, and accurate.
|
|
|